Clamp diodes are widely used in analog electronic
circuits, but, like all real-world components, these diodes have some non-ideal characteristics. This article investigates some of their
characteristics.
Reverse Biased Leakage Current
Reverse biased leakage current is a characteristic of
concern for a clamp diode. Ideally the leakage would be zero but this is the
real world, so what can one live with? In a typical application there might be
two clamp diodes, one to +15 V the other to -15 V and a common connection to
the signal to be clamped. The leakage current drops as the voltage drops, so if
that signal is nearer to +10 V, then the lower clamp diode would have more
leakage than the upper one. So in such an application, a diode with both low
leakage and a fairly flat leakage vs. voltage would be desirable.
with an AMETRIX Instruments’ Model 101 Picoammeter to
measure the current (and yes, that is single digit femtoamps in the diagram).
Comments about reverse leakage:
1. The SST4391 is rated for 35 V gate-drain and
gate-source
2. The SST4391 costs $0.057 each
3. The 2n3904 has a Vcb maximum of 60 V
4. The MMBT3904LT1G is $0.0231 each
5. The FLLD261 is $0.0275 each
6. The BAS16 costs $0.0151 each
7. The BAT54 costs $0.0204 each
8. The BAS70 costs $0.0323 each
9. The MMBD4148CC costs $0.0194 each
So even though there is a 3:1 price ratio between the most
and least expensive, they are all <6 a="" cost="" decision="" is="" not="" of="" part="" process.="" serious="" so="" span="" the="">6>
Forward Voltage Drop
The forward voltage drop is another diode characteristic
to be considered for a voltage clamp. These measurements were made as follows
(the data is presented graphically in Figure 2):
1. Forward bias the diode with a quiet DC supply that has
fine resolution; in this case an EDC calibrator
2. With function generator set to pulse, 1 Hz, 1% duty
cycle, drive a mercury wetted reed relay. Connect
the relay contacts across DUT such that the DUT is shorted
for 990 msec and unshorted for 10 msec; this is to minimize self heating
effects.
3. Measure forward voltage with Tektronix MSO4104 scope
with 10x probes to minimize the low forward current errors. Set the scope’s
bandwidth limit to 20 MHz to minimize noise. Use the scope’s vertical channel’s
offset mode and 10 mV range, adjusting the offset knob until trace is in center
of screen, then document the offset voltage.
Conclusion
Some simple measurements with a precision picoammeter can provide much
insight into the components that you think might be ideal in a given application; and they might
reveal some things that are unexpected.